INFICON SKY CDGxxxD (EtherCAT) User Manual
Page 18
18
tira68e1 (2013-05) CDGxxxD.cp
High Trip Point
Time t
in-
active
active
in-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
Index
SI
DataType
NV Access PM Name
800E
0x01 BOOL
x
RO
High Trip Enable
0x02 BOOL
x
RO
Low Trip Enable
0x11 REAL
x
RO
High Trip Point Limit
0x13 REAL
x
RW
Percentage High Trip
Source
0x14 REAL
x
RW
Low Trip Point Limit
0x17 REAL
x
RO
High Trip Hysteresis
0x18 REAL
x
RW
Low Trip Hysteresis
0x1A
UDINT
x
RO
Source
Index
High Trip Point
0
1
Disable
Enable
Low Trip Point
0
1
Disable
Enable
High Trip Point Limit: High limit to trigger trip point condition if Input Value (I
0x900E, SI 0x01) is above this limit.
Percentage of High Trip Source: Percentage of Piezo Sensor Value (I 0x6010, SI
0x11) in [%].
Low Trip Point Limit: Low limit to trigger trip point condition if Input Value (I 0x900E,
SI 0x01) is below this limit.
High Trip Hysteresis: Hysteresis value for High Trip Point.
Low Trip Hysteresis: Hysteresis value for Low Trip Point.
Subindex 0x01
Subindex 0x02
Subindex 0x11
Subindex 0x13
Subindex 0x14
Subindex 0x17
Subindex 0x18