JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual

Page 22

Advertising
background image

Preparations

22

Backflow safety device

T16A

T16A

T1.25A

ext

Pt100

SERIAL

ALARM

REG+
E-PROG

STAND-BY

13

14

(see page 66

If the liquid levels in the circulator bath and the external
system are at different heights, overflowing must be
prevented after the power has been turned off.

Flood hazard!

For this reason, solenoid valves for loop circuit or shut-off
valves can be integrated in the loop circuit.

Connect the valve to the connector (13).

If socket (14) is used:
In menu >Inputs/Outputs< set the
menu item >SV-Out< to >Backflow<.

Order No. Description
8 980 701 Set of solenoid valves

for loop circuit up to +100 °C

8 970 456 Shut-off valve (suitable up to +90 °C)
8 970 457 Shut-off valve (suitable up to +200 °C)

5.3.1. Tubing

Recommended tubing:

Order No.

Length

Temperature range

8930008 1

m

CR

®

tubing

8 mm inner dia.

-20 °C to 120 °C

8930012 1

m

CR

®

tubing

12 mm inner dia.

-20 °C to 120 °C

8930108

1 m

Viton tubing

8 mm inner dia.

-50 °C to 200 °C

8930112

1 m

Viton tubing

12 mm inner dia.

-50 °C to 200 °C

8930410

1 m

Insulation for tubing 8 mm inner dia.

-50 °C to 100 °C

8930412

1 m

Insulation for tubing 12 mm inner dia.

-50 °C to 100 °C

8 930 209

0.5 m

8 930 210

1.0 m

8 930 211

1.5 m

8 930 214

3.0 m

Metal tubing, triple insulated, M16x1

-100 °C to +350 °C

8 930 220

0.5 m

8 930 221

1.0 m

8 930 222

1.5 m

8 930 223

3.0 m

Metal tubing, insulated, M16x1

-50 °C to +200 °C


Warning:

Tubing:

At high working temperatures the tubing used for temperature application and cooling
water supply represents a danger source.
A damaged tubing line may cause hot bath fluid to be pumped out within a short time.
This may result in:
 Burning of skin

 Difficulties in breathing due to hot atmosphere

Advertising