JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual

Page 27

Advertising
background image

FP51-SL

27

Warning:

The excess temperature protection m u s t be set at least 25 °C below the fire point of
the bath fluid used!
In case of wrong setting there is a fire hazard!
We disclaim all liability for damage caused by wrong settings!

SafeTemp:

Setting range: 20 °C ... 320 °C

Rough setting can be effected by using the
temperature scale.

Settings:

 Press the key

and by pressing the key

or

select

the menu item

and confirm the by pressing the key

.

For setting proceed as follows.

SafeTemp:

 Set the new cut-out value within 30 seconds with a screw driver.

Exact display and setting on LCD display
Recommendation:
Set over temperature-protective setting >SafeTemp< 5 °C to 10 °C
above the set point of the working temperature.

Type:

Switch-over from warning to shutdown function

 Change the parameter by pressing the key

and confirm by

pressing the key

.

or

pressing the key

if the parameter is to be retained.

Possible parameters: warning or alarm

OverTemp :

SubTemp:

 Set value on numeric keypad and confirm by pressing


Example: 83.00 °C

Press the keys

and

and confirm by pressing

SafeSens:

Indicated is the temperature value of the safety sensor.



SafeTemperature:

Setting range: 20 °C ... 320 °C

Indicated is the cut-out value set with a screwdriver on the excess
temperature protection device.

This safety installation is independent of the control circuit. When the
temperature of the bath fluid has reached the safety temperature, a
complete shutdown of the heater and pump is effected.

The alarm is indicated by optical and audible signals (continuous tone).

Advertising