52

9.5.1.

Control internal / external

The control type can be adjusted in the -OFF- mode only.

Possible parameters:

intern

internal temperature control

extern

external temperature control with external Pt100 sensor

ext.

Pt100

IMPORTANT:
Additional measures for external temperature control.

Connect a Pt100 sensor to the socket on the rear of the circulator.

Sensor calibration of the external Pt100 sensor is performed in the
>ATCalibration< menu.

Suggested adjustments for external temperature control:
Upper band / Lower band and IntMax / IntMin
see Section >Limits<.

External temperature control does not allow the setpoint to be set via
the “ext. Pt100” socket.

Notice:

Place the external sensor into the temperature-controlled medium and securely fix the
sensor.


Pt100 M+R

Accessory: Pt100 external sensor

Order No.

Description

Material

Cable

8981003

200x6 mm Ø,

stainless steel

1.5 m

8981005

200x6 mm Ø,

glass

1.5 m

8981006

20x2 mm Ø,

stainless steel

1.5 m

8981010

300x6 mm Ø,

stainless steel

1.5 m

8981015

300x6 mm Ø,

stainless steel / PTFE coated

3 m

8981013

600x6 mm Ø,

stainless steel / PTFE coated

3 m

8981016

900x6 mm Ø,

stainless steel / PTFE coated

3 m

8981014

1200x6 mm Ø,

stainless steel / PTFE coated

3 m

8981103 Extension cable for Pt100 sensor

3.5 m

8981020 M+R in-line Pt100 sensor

The M+R in-line Pt100 sensor is a flow sensor and can be installed loop
circuit