9.7. ATC Absolute Temperature Calibration, 3-point calibration

ATC is used to compensate for a temperature difference between the
circulator and a defined measuring point in the bath tank that may
develop for physical reasons.
ATC

internal

external



Example:
1-point calibration

°C

°C

T 1

T


3-point calibration

°C

°C

T 1

T

T

T

1 = Original curve

Principle:
For ATC calibration, in steady state the bath temperature at the location
of the temperature sensor (CT) is determined at the respective adjusted
working temperature. This value is then set on the circulator in the menu
>ATCalibration< under menu item > CTEMP X <.
This can be a 1-point, 2-point or 3-point calibration.

°C

TT

M

B (INT)

T

ATC (internal)

°C

TT

M

B (EXT)

T

Pt100

ext

ATC (external)


M = Temperature measuring instrument with temperature sensor
B = Bath tank (INTernal or EXTernal)
T = circulator

CT = Temperature on measuring point