12 3 ak – JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual

Page 80

Advertising
background image

Electrical connections

80

RS232 interface cable
Circulator

(9-pol)

PC

(9-pol)

Pin 2 RxD

Pin 3 TxD

Pin 3 TxD

Pin 2 RxD

Pin 5 GND

Pin 5 GND

Pin 7 RTS

Pin 8 CTS

Pin 8 CTS

Pin 7 RTS

Accessories: Order

No.

Description

8 980 073

RS232 interface cable 9-pol./9-pol. , 2,5 m

8 900 110

USB interface adapter cable



/ Control output

The connector may only be used for control of a JULABO refrigerated
circulator or JULABO MVS Solenoid valve controller for cooling water.

Pin assignment:
Pin

Signal (only with attached JULABO equipment)

1

+24 V (I max. current 25 mA)

2

0 V

3

Alarm relay

4

Reserved - do not use!

5 Cooling

pulse

ALARM

Alarm output

(for external alarm signal)

This potential-free change-over contact is activated in case of an alarm
when pins 2 and 3 are connected.

Switching capacity

max. 30 W / 40 VA

Switching voltage

max. 125 V

/

Switching current

max. 1 A

1

2

3

AK

STAND-BY

Stand-by input
(for external switch-off)

Pin assignment:

Pin

Signal

1

not

connected

2

5 V / DC

3

0 V

Advertising