9.9.3.

Input of the connector - REG+E-PROG



Menu >Configuration<

1.

2.

Menu >Inputs/Outputs<

3.

E-PROG - Input

Setting is necessary if

1. the Setpoint is to be set via an external voltage or current source or

programmer
For this, in the menu >Configuration<, first set the menu item
>Setpoint< to >Eprog<.

2. the heater variable should be controlled via an external control

pulse.
For this, in the menu >Configuration<, set the menu item >
ActVar.< to >Eprog<.

3. the signal of an external flow meter should be registered.

The E-Prog input can only be used either under menu item
>Setpoint< or under menu item > Actvar <.
If the input is occupied neither by >Setpoint< or > ActVar <, the
signal of a flow meter can be connected.

First set menu >Configuration<, then menu >Inputs/Outputs<.









Example:

Connect the external voltage or current source or programmer to
the REG+E-PROG socket (12) of the circulator.

Selecting the signal:

The programmer (E-PROG) input of the circulator can be matched
to the output signal of the external voltage or current source.
Voltage

voltage

input

Current

current

input

“L Value” - Setting the LOW value: (See below

)

First adjust and set the lowest voltage or current on the external
voltage or current source (e.g., 0 V or 0 mA).
Then after approx. 30 seconds, enter the corresponding
temperature value (e.g., 20.00°C).on the circulator by pressing the

appropriate keys on the keypad and press

to set.

“H Value” - Setting the HIGH value: (See below

)

First adjust and set the highest voltage or current on the external
voltage or current source (e.g., 10 V or 20 mA).
Then after approx. 30 seconds, enter the corresponding
temperature value (e.g., 300 °C).on the circulator by pressing the

appropriate keys on the keypad and press

to set.


Example:

Set the external voltage or current source output for the equivalent
of a 50 °C temperature setpoint.
The value adjusted and set on the external programmer is
displayed in line 4 of the LCD DIALOG-DISPLAY for inspection.