List of commands – JULABO FP51-SL Ultra-Low Refrigerated Circulators User Manual

Page 83

Advertising
background image

FP51-SL

83

Important times for a command transmission:

To ensure secure data transfer, the time gap between two commands
should be at least 250 ms.
The circulator automatically responds to an in command with a data string
followed by an LF (Line Feed). The next command should be sent only
after 10 ms.

The out commands are valid only in remote control mode.

Examples:

When the RS485 interface is used, the three-digit instrument address
precedes each command.
(example: address Ad32 = A0 3 2 )

Command to set the working temperature >Setpoint1< to 55.5 °C

out_sp_00

55.5

A032_out_sp_00

55.5

Command to retrieve the working temperature >Setpoint1<

in_sp_00

A032_in_sp_00

Response from the circulator:

55.5

LF

A032_55.5

LF

Notice:

out commands

Temperature values can be displayed on the circulator in °C or °F.
The settings sent via the interface must also be made in the selected unit according
to this setting.


12.3. List of commands

out commands: Setting parameters or temperature values.

Command

Parameter

Response of circulator

out_mode_01

0

Use working temperature >Setpoint1<

out_mode_01

1

Use working temperature >Setpoint2<

out_mode_01

2

Use working temperature >Setpoint3<

out_mode_02

0

Selftuning “off”.
Temperature control using the stored parameters.

out_mode_02 1

Selftuning

“once”

Single selftuning of the controlled system after the next start.

out_mode_02 2

Selftuning

“always”

Continual selftuning of the controlled system whenever a new
setpoint is to be reached.

out_mode_03

0

Set external programmer input to voltage.

Voltage

0 V ... 10 V

Advertising