8 Configuration
8.5 Decimal point (decimal place)
DP
Select the position of the decimal point.
Select between:
0.1 = one decimal place and
1 = no decimal place
factory-set:
0.1
h Press the UP ( ) or DOWN ( ) key to change between the formats.
h Press the SET key to activate the selected format.
h Press the ESC key twice to return to normal display.
8.6 Temperature correction
PV Bias
Using this parameter, the temperature can be corrected by a constant value,
e.g. with known measurement errors.
Select between:
-99.9 to +99.9 °C (°F)
factory-set:
0.0
h Select the digit to be altered by pressing the LEFT ( ) or RIGHT ( ) key
(digit is shown blinking).
h Press the UP ( ) or DOWN ( ) key to alter the value.
h Press the SET key to activate the selected correction.
h Press the ESC key twice to return to normal display.
Example
“PV Bias” is set to 1.0°C. The temperature is 99.0°C. The sum of both values
is displayed and stored (100.0°C).
blinking
blinking
Measured
temperature
99.0°C
PV Bias
1.0°C