2 problems and solutions, 4 technical data, 2problems and solutions – JUMO 702540 TDA-300 / TDA-3000 - Handheld thermometer Operating Manual User Manual
Page 66: 10 appendix
10 Appendix
66
10.3.2 Problems and solutions
10.4 Technical data
v See Data Sheet 702540
Problem
Possible cause
Solution
Instrument cannot be
switched on (no
display)
No battery fitted or
battery is discharged or
voltage too low.
Measured
temperature is
faulty
Battery voltage too low.
Interference from
another device.
- Remove the device that
causes the interference.
- Use the parameter
“digital filter”.
Wrong sensor is used.
- Use the correct sensor.
Sensor connector not
inserted far enough.
- Check the sensor
connection to
instrument.
Measured
temperature is not
identical with the
actual temperature
Temperature correction
(PV bias) is used.
- Switch off temperature
correction.
Display does not
respond, instrument
does not react to a
key stroke.
CPU overrun
- Remove the
battery, wait a
few seconds, then re-
insert it.
If the error does not
disappear,
contact the supplier.