10.3.2 Problems and solutions

10.4 Technical data

v See Data Sheet 702540

Problem

Possible cause

Solution

Instrument cannot be
switched on (no
display)

No battery fitted or
battery is discharged or
voltage too low.

v Chapter 2.1 “Inserting

the battery”

v Chapter 10.1

“Replacing the battery”

Measured
temperature is
faulty

Battery voltage too low.

v Chapter 10.1

“Replacing the battery”

Interference from
another device.

- Remove the device that

causes the interference.

- Use the parameter

“digital filter”.

v See “D filter” on Page

50.

Wrong sensor is used.

- Use the correct sensor.

Sensor connector not
inserted far enough.

- Check the sensor

connection to
instrument.

Measured
temperature is not
identical with the
actual temperature

Temperature correction
(PV bias) is used.

- Switch off temperature

correction.

v See “PV Bias” on Page

49.

Display does not
respond, instrument
does not react to a
key stroke.

CPU overrun

- Remove the

battery, wait a
few seconds, then re-
insert it.

If the error does not
disappear,
contact the supplier.