Adagio User Manual

© 2010 KD Scientific

5619-001 Rev B

35

The number of doses, dose value and lag time fields are available only if
the “Multiple Doses” option is selected. This means that a step sequence
is to be built from a known dose and concentration.


The duration value field is available only if the “Single Dose” option is
selected. This means that a single step is to be built from a known
concentration and duration.


All numerical values (except the doses count, which must be an integer
value) can be entered with up to 4 decimals making use of the decimal
symbol currently configured in the Microsoft® Windows® regional
settings.

The step sequence is generated from the data entered in the wizard and
depends mainly on the option selected:

Multiple Doses: a step sequence is built to repeatedly infuse /

withdraw a volume of :

V (ml) = ([Dose to apply (mg/kg)] * [Subject weight (kg)]) /

[Concentration (mg/ml)]


as many times as [Number of doses] states with a break time of [Lag

time (min]] between two consecutive doses.

That means that one step is added for each dose with the following
parameters:

o

Duration (min) = V (ml) / [Flow rate (ml/min)]

o

Start flow (ml/min) = [Flow rate (ml/min)]

o

End flow (ml/min) = Start flow

o

Step linkage property = Not linked



Each of these steps is then followed by the corresponding lag step
with the following parameters:

o

Duration (min) = [Lag time (min)]

o

Start flow (ml/min) = 0

o

End flow (ml/min) = 0

o

Step linkage property = Not linked