Instrument setup – Leica Biosystems SM2010 R User Manual

Page 17

Advertising
background image

15

Leica SM2010 R

Fig. 13

Setting the clearance angle (

Fig. 13

)

• Unscrew the knurled head screw (69).
• Set the required angle with the adjusting lever

(70) using the scale for the clearance angle

(71). The upper edge (73, red arrow in

Fig. 13

)

of the lever (70) has to align with the index

mark of the selected degree value.

• To fix the setting in place, retighten the knurled

screw (69).

The clearance angle can also be adjusted if a

knife is clamped.
• To do this, slightly unscrew the two clamping

screws for the knife (72) and, if necessary, also

the knurled head screw (69).

• Set the desired angle as described above.
• To fix the setting in place, retighten the knurled

screw (69).

Enlarged detail:

Scale for

setting the

clearance

angle.

72

69

70

73

4. Instrument Setup

Inclination (declination)

of the knife holder SN
• Loosen the clamping lever (66).

• Set the required declination (inclination of the

knife holder from the cutting direction) on the

scale (55.1) at the back of the knife holder (67).

• On the intermediate plate (60) there is an index

mark (68.2) which serves as a point of refer-

ence for the scale division when adjusting the

declination.

• For fixation, tighten the clamping lever (66)

firmly in the required position.

66

67

60

68.1

68.2

Fig. 14

Assembling the knife holder SN

(continued)

71

70