Leica SM2010 R

5. Operation

5.3 Clamping a disposable blade

Insert the blade (Fig. 28)
• Push the knife guard (14) towards the right

and push the lever (46) upwards to release

the clamp of the pressure plate (48).

• Flap the insertion aid (57) downwards.

• Position the dispenser (4) with the dispos-

able blades sideways (see Fig.) and push the

blade (6) into the blade holder.

• Using a brush stick (47), carefully push the

blade into its final position. The notch (49)

makes it easier to push the blade all the way

underneath the pressure plate.

• Push the clamping lever (46) downwards to

clamp the blade.

• Make sure that the blade is clamped parallel

to the front edge of the pressure plate.

Removal aid (Fig. 29)
A removal aid is provided for removing used

blades (6).

• Loosen the clamping lever (46) of the pressure

plate (48).

• Push the knife guard (14) towards the left while

holding down the black push button (51). This

pushes out the blade far enough sideways that

it can be removed easily.

48

6

14

46

47

49

57

4

46

51

6

51

Fig. 29

48

Fig. 28

Be very careful when handling micro-

tome knives or blades. The cutting edge

is extremely sharp and can cause seri-

ous injuries!

The blade holder has to be installed in

the instrument before a blade is inserted!