Instructions for Use V 1.4 RevC – 09/2013

5.4 Replacing the pressure plate

The blade holder SE can be converted

from use of narrow blades to use of

wide blades, and vice versa.

To do this, both the pressure plate (48)

and the matching insertion aid (57)

have to be replaced.

To replace them, proceed as follows:

• Push the knife guard (14) towards

the right and push the lever (46)

upwards to release the clamp of

the pressure plate (48).

• Now carefully pull the insertion

aid (57) out to the left (

Fig. 30

). The

pressure plate (48) can now be

taken off (

Fig. 31

).

• To mount another pressure plate,

please proceed in the reverse

sequence. Only use the pressure

plate together with the matching

insertion aid.

21

Fig. 32

Fig. 30

Fig. 31

48

46

14

57

Pressure plate

kit

48

57

Insertion aid

Pressure plate

21

5. Operation

The slotted screw (21) in

the pressure plate is ad-

justed at the factory and

glued in. It must not be

altered.