5.1.3 Manual feed

The lever (35) for the manual feed is at the front

right of the instrument.
• Each time the lever is pushed or pulled, this

causes a feed motion for trimming or section-

ing by the value set on the scaled adjusting

knob (33).

5.1.4 Automatic feed

The position of the adjusting knob (22) determines

the point of the knife sledge movement where the

automatic feed takes place.

It should take place immediately in front of the

specimen.

33

35

5

6

22

22

23

Fig. 24

Fig. 25

If the adjusting button (22) is locked in

the frontmost (left) position (

Fig. 25

),

there is no feed motion.

The automatic feed is only effective up

to a section thickness of 30 µm. All val-

ues set beyond that are not defined.

• To adjust the automatic feed, move the blade/

knife (6) until it is just in front of the specimen

(5) (

Fig. 25

)

and lock it in this position using the

locking knob (23).

• Loosen the adjusting button (22) and push it

back until you can feel resistance. Retighten

it in this position.

• For exact work with the automatic feed, the

knife sledge must be moved past the position

of the adjusting knob.