5 directional fixture for specimen clamps, Operation – Leica Biosystems SM2010 R User Manual

Page 25

Advertising
background image

23

Leica SM2010 R

5. Operation

5.1.5 Directional fixture for specimen clamps

Orienting the specimen
• To loosen the clamp, rotate the eccentric

lever (34) upwards (until you feel it click into

position).

Caution! Turning it further to the left re-clamps

the orientation!

• Turn setscrew (30) to orient the specimen in

the cutting direction. Turn setscrew (31) to

orient the specimen transverse to the cutting

direction.

Each complete turn of the screw inclines the

specimen by 2°. A total of 4 complete turns =

8° are possible in every direction.

The accuracy is approximately ± 0.5°.

For better orientation, there is a noticeable

notch point after each complete turn of the

setscrew.

• To lock the current orientation, turn the ec-

centric lever (34) backwards.

Display of the zero position
For better display of the zero position, each set-

screw (30, 31) has a red mark (32).
When both marks are visible and both setscrews

are in zero position at the same time (notch point!),

the specimen orientation is in zero position (0°).

The object orientation allows for simple position

correction of the specimen surface when the

specimen is clamped into place.

32

30

34

31

29

All object clamps available as optional accessories can be inserted into the quick

clamping device (29) of the directional specimen holder fixture, all object clamps avail-

able as optional accessories can be used.

Fig. 26

Advertising