16

Operating Manual V 1.0 – 01/2010

Fig. 8

4. Setup the Instrument

4.5 Connecting the foot switch

The standard scope of delivery of the "Profes-

sional" and "Professional Gold" instrument ver-

sions include a foot switch (18,

Fig. 8

) that can

be connected to the rear side of the instrument.
In SINGLE and AUTO modes, the sectioning pro-

cess can be started and stopped with the foot

switch instead of the DIRECTION switch.

To connect the foot switch, follow these steps:

• Route the cable of the foot switch so that the

switch on the floor can be reached easily

with one foot.

• Insert the plug (17) of the foot switch into

the socket (19) provided on the rear side of

the instrument. This is possible in one posi-

tion only; it is not possible to connect it the

wrong way.

• Then screw the plug and the retaining ring

(20) securely onto the thread of the socket

(19).

• Before the foot switch can be used, the small

switch (21) must be moved into the "ON" po-

sition, (Fig. 13).

If the foot switch is installed and "ON",

the sectioning process cannot be

started or stopped with "DIRECTION".

17

18

19

20

Rear side of the

instrument

21