Operating Manual V 1.0 – 01/2010

6.3 Replacing components
6.3.1 Fuse

If the instrument fails to function completely, first check for power avail-

ability at the power outlet. The instrument fuse(s) located at the rear of

the instrument should be checked next. Before removing the fuse, pull the

POWER PLUG OUT OF THE SOCKET. The fuse and its holder are removed

by lifting the tab in the fuse drawer that is located directly above the plug

and pulling the drawer out of the socket.
Check to see if the filament within the glass tube of the fuse is intact. If

not, replace with fuse referenced below, two of which are supplied in

the accessory package. Before reconnecting the power cord, check the

instrument for any obvious cause of the burnt fuse and make the appropri-

ate correction.

Model #

Qty

Fuse

064018

1

GMA-2A

064026

2

F-1A

6.3.2 Fluorescent lamp

If the fluorescent lamp fails to turn on or flickers when operating, replace-

ment is necessary. The fluorescent lamp is located under the lamp assem-

bly housing, directly behind the glass magnifier.

BE ABSOLUTELY CERTAIN TO UNPLUG THE POWER CABLE before

removing the lamp.

The lamp can be removed from its mounting sockets by first removing the

end cap of the lamp housing opposite to the socket. Once the socket is

removed, the lamp tube can be removed by pulling the lamp straight out of

the socket.