Operation – Leica Biosystems VT1000 P User Manual

Page 25

Advertising
background image

25

Leica VT1000 P

5. Operation

The operating elements

(continued)

Fig. 32

AMPLITUDE rotary knob

This knob controls the relative amplitude of the lateral excursion of

the sectioning blade vibratory movements.
• In the zero position (‘0‘) or during the retraction movement of the

cutting head, the blade does not vibrate.

• The maximum amplitude (2.0 mm) is reached when the rotary knob

is in the ‘10‘ position. At intermediate settings, the amplitude will

vary proportionally.

All configured values are active during the feed movement of

the cutting head only.

The amplitude can also be changed during the sectioning pro-

cess.

LIGHT switch

Fig. 33

This simple switch switches on the fluorescent lamp and thus im-

proves the illumination of the specimen tray. Switch position:
ON --> Light is on.
OFF --> Light is off.

Advertising