Setup the instrument – Leica Biosystems VT1000 P User Manual

Page 19

Advertising
background image

19

Leica VT1000 P

4. Setup the Instrument

Fig. 17

4.7 Assembling the specimen clamp

• The specimen clamp (34,

Fig. 16

) is assem-

bled inside the regular black Vibratome tray

(23,

Fig. 15

).

• Insert the cheese head screw (35) into the

bore of the specimen stage (27) and tighten

it using a

size 5

Allen key (37, included in the

scope of delivery).

• When tightening, ensure that the fixture (34)

is parallel to the front edge of the Vibratome

tray (

Fig. 17

).

• Insert the specimen block (36) into the clamp

and clamp it into place using the knurled

screw (39) (

Fig. 18)

.

• The specimen to be sectioned is affixed to

the specimen block using the cyanoacrylate

adhesive provided.

• The purpose of the knurled screw (38) is to

ensure that the

directional specimen fixture

can also be installed (for additional informa-

tion, refer to

Fig. 22

).

Fig. 16

23

27

34

Fig. 15

35

36

37

37

34

Fig. 18

36

38

39

Always remove the blade BEFORE in-

stalling or removing accessories!