Leica VT1000 P

5.2.3 Adjustable sectioning window

The VT1000 A features an adjustable sectioning window that you can use

you speed up the process of making series sections.

This "window" is the distance the blade travels

from a resting retracted position to the front of

the bath when it automatically ends the forward

stroke and reverses.
This corresponds to distance "A" in

Fig. 46

.

Fig. 40

If the front and rear edge of the "window" are

positioned as close as possible to the front and

rear end of the specimen, it is only necessary to

cover distance "B" for a section.
This reduces the time for a complete cycle or

section significantly.