6. Cleaning and Maintenance

6.1 Cleaning the instrument

• Always remove the blade before detaching the blade holder from

the instrument!

• Always keep the sapphire blades in the box when not in use!

• Never place used blades on the laboratory table - dispose of them

safely!

• When using detergents, comply with the safety instructions of the

manufacturer and observe the laboratory regulations valid in the

country of use.

• When cleaning the outer surfaces, do not use xylene, scouring

powders or solvents containing acetone or xylene. The finished

surfaces are not resistant to xylene or acetone!

• Ensure that no liquids enter the interior of the instrument when

cleaning!

Before each cleaning, carry out the following preparatory steps:

• Move the specimen clamp into the lower end position.
• Remove the blade from the blade holder and dispose of it safely (be-

fore disposing of used blades, apply thick "Scotch" tape to the cutting

edge or wrap the entire blade in paper) or place the sapphire blade in

the knife case.

• Remove the specimen from the specimen clamp.
• Remove section waste and dispose of it safely.
• Remove specimen clamp and clean separately.

• Do not open any covers or casing panels of the instrument, as

electrically live parts are on the inside!