Operation – Leica Biosystems VT1000 P User Manual

Page 24

Advertising
background image

24

Operating Manual V 1.0 – 01/2010

5.

Operation

The operating elements

(continued)

This switch and its three switch positions (upper, center, lower) con-

trol the movement direction of the cutting head.
• Push the switch upwards to start the sectioning cycle. The sec-

tioning blade will advance forward at maximum speed when the

switch is held at its top position.

• As soon as the switch is released, it returns to the center position.

• The blade continues the feed movement with the default values for

SPEED and AMPLITUDE.

• At the end of the sectioning window, the movement direction of

the blade is reversed and the cutting head moves back. Reversing

stops when the rearmost "ready" position is reached.

• During the feed movement, the movement direction can be

reversed at any time by moving the switch to the lower position.

• Likewise, during the retraction movement, you can switch to infeed

by pushing the switch upwards.

DIRECTION switch

Switch position:
Upwards --> forwards

Downwards --> backwards

Rotary knob

SPEED

Using this rotary knob, you can change the relative speed of the cut-

ting head (feed movement) if the DIRECTION switch is in the

FWD/

AUTO REV

position.

• In the zero position (‘0‘), the blade does not move; in the ‘10‘ posi-

tion, it is moved at maximum speed (2 mm/s).

• For the positions in between, the speed is changed proportionally.

Fig. 30

Fig. 31

Only the feed rate can be changed; this is also possible during

the sectioning process.

The retraction speed is always the same.

Advertising