Instrument components and specifications – Leica Biosystems EG1150 C User Manual

Page 11

Advertising
background image

11

Leica EG1150 C

3.2 Technical data

General data

Approvals:

The instrument-specific marks are located on the rear

panel of the instrument next to the name plate

The 120 V version has the test symbol c-CSA-us.

Voltage of power supply:

The device requires a specific AC volt-

age, see table at right.

Instrument type acc. to identification label.

AC ±10 %

Power fuse:

T1A L250 V

Nominal power:

350 VA

Maximum start-up current (5 s)

25 A

Operating temperature range:

+15 °C to +28 °C

Operating temperatures:

-5 °C up to a room temperature of approx. +28 °C

Relative humidity:

20 to 80 % - non-condensing

Permissible temperature range

during storage and transport:

-29 °C to +50 °C

Permissible humidity range

during storage and transport:

10 to 85 % - non-condensing

IEC 1010 classification:

Protection class 1

Pollution degree

2

Operating elevation:

up to max. 2000 m NN

IP protection class (IEC 60529):

IP20

Refrigeration unit

Refrigeration capacity*:

110 W

Safety factor:

3

Refrigerant:

R 134a

Compressor oil:

180 cm

3

Emkarate RL 15 s, ICI

*according to CECOMAF, fluid temperature: 55 °C, evaporation temperature: -25 °C

Dimensions and weight

Width:

360 mm

Depth:

650 mm

Height:

360 mm

Weight:

23 kg

Instrument type Power supply voltage

Coolant

EG1150 C-1

230 V / 50 Hz

80 g ± 5 g

EG1150 C-3

120 V / 60 Hz

80 g ± 5 g

EG1150 C-8

240 V / 50 Hz

80 g ± 5 g

EG1150 C-11

100 V / 50 - 60 Hz

65 g ± 5 g

3. Instrument Components and Specifications