16

Leica EG1150 C

5. Operation

A miniature fuse to protect the electronic com-

ponents is located on the rear of the instrument.

To replace the fuse, please proceed as follows:
• Using a screwdriver, lightly press the fuse

holder (9) and turn counterclockwise about

1

/

4

turn.

• Remove the fuse holder and take out the

fuse (10).

• Replace it with a new fuse of the same type.

• Replace the fuse holder (9).

• Using a screwdriver, gently press the fuse

holder and turn it clockwise about

1

/

4

turn.

• Reconnect the instrument to an AC power

outlet and switch it on.

Fig. 7

5.1 Switch on the instrument

After installation as described in

Chapter 4

, the Leica EG1150 C is ready for

operation. Switch on the device with the On-Off switch at the front left of

the instrument ("I" = ON). A lamp lights up in the switch to indicate that the

unit is operational and the refrigeration unit will start working.

Depending on the room temperature, the working temperature of the cool-

ing surface (-5 °C) will be reached in around 25 minutes (see also

Ch. 4.1

).

9

5.2 Replacing the secondary fuse

10

9

Before replacing the fuse, always

switch the instrument off and pull the

power plug from the wall socket.

Only miniature fuses of the type speci-

fied in the chapter "Technical data"

may be used.

The cooling surface may not be loaded with molds until the cooling

time has elapsed. Otherwise, the working temperature of -5 °C may

not be reached.