5. Operation

5.3 Instrument malfunctions

If you cannot solve your problem using the help in the following table, please contact your Leica

customer service organization or the Leica dealer from whom you purchased the instrument.

Malfunction

Possible causes

Corrective action

Instrument doesn't work.

Power switch not turned on.

or

The cut-out function of the

power switch activated.

Check instrument is connected,

then turn main power switch

on.

Only for Great Britain:

Check the fuse in the power

plug.

Cooling unsatisfactory,

formation of condensation.

Room temperature too high or

exposure to direct sunlight.

Check the cooling again at a

lower room temperature.

Inadequate air supply to

ventilation unit.

Check distance of instrument to

wall; it must be at least 15 cm.

Refrigeration system leak.

The instrument must be

serviced or repaired by

a service technician.

Please contact customer

service.

Compressor switches off after

a brief period of operation

and switches back on shortly

thereafter.

AC power does not match

specifications or

Compressor defective.

Check the voltage of power

supply.

Compressor does not start.

Compressor defective.

The instrument must be

serviced or repaired by

a service technician.

Please contact customer

service.

Compressor switches off after

a short period of operation.

Compressor defective.