2.3 Integrated safety devices

Automatic circuit breaker in the main switch
• An automatic circuit breaker with a lit

switch is located in the main switch. This

circuit breaker disconnects the unit from the

power supply in the event of a short circuit.

• In this case, the main switch jumps to the

position "0" = Off.

Fig. 2

Short-circuit

2. Safety

• The minimum voltage (see Technical data) must be maintained while starting the refrigera-

tion unit.

The compressor needs a start-up current of approx. 25

A. A stable power supply in accor-

dance with the instrument's specifications is essential to its proper functioning. Please

ensure that your electrical installation fulfills these preconditions prior to installing the unit.

Nonobservance causes damage to the instrument.

• Switch off the instrument each time before servicing, repairing or cleaning, and pull out the

power plug.