Leica EG1150 C

4. Instrument Setup

4.4 Power supply

The Leica EG1150 C refrigeration unit requires a specific voltage and

frequency (refer to

"Technical data" chapter

), and is therefore always

delivered with a power cable that fits the instrument.
Please observe the following notes to prevent damage to the unit:

• Connect the power cord plug (7) to the connecting port (6).
• Plug the power cord into the wall outlet.

Fig. 6

6

7

Only one of the following network cables may

be used to connect the instrument:

Application area

Order No.

Australia

14 0411 32565

Europe:

14 0411 13558

Italy:

14 0411 13561

Switzerland:

14 0411 13560

United Kingdom:

14 0411 27822

USA, Canada, Japan:

14 0411 13559

Brazil

14 0411 47869

Before connecting the instrument to the power supply, it is manda-

tory to check whether the voltage specified on the identification la-

bel (rear side) matches the actual voltage values at the installation

location.

If this is not the case, the connection must not be made!

The unit must be connected only with the supplied power cable and

only to a grounded power receptacle.

Do not use an extension cord!