Model 409.83 temperature controller specifications – MTS Model 409.83 Temperature Controller User Manual

Page 16

Advertising
background image

Model 409.83 Temperature Controller

16

Model 409.83 Temperature Controller Specifications

Introduction

interlock when the heater elements are enabled. If undesirable conditions arise,
power to the heater elements is cut off automatically.

The following figure illustrates a block diagram of the Model 409.83
Temperature Controller and a high-temperature furnace.

A temperature setpoint is selected for each furnace heat zone. The setpoint is
compared to the temperature measurement provided by a thermocouple inside in
the furnace. Any difference (between the setpoint and the actual temperature)
causes an error signal.

The error signal produces a time proportional signal. This time proportional
signal is sent to a corresponding solid state relay. The time proportional output
pulses the solid state relay on and off at a rate determined by the amplitude and
polarity of the error signal. The solid state relay applies the AC power to the
corresponding heating element in the furnace.

Model 409.83 Temperature Controller Specifications

High
Temperature
Furnace

Upper Control
Signal

Middle Control
Signal

Lower Control
Signal

Lower Thermocouple

Signal

Middle Thermocouple Signal

Upper Thermocouple Signal

element

element

element

Temperature
Controller

Parameter

Specification

*

Environmental:

Temperature

Relative Humidity

0°C to 50°C (32°F to 122°F)

10 to 90%, noncondensing

Line Voltage

100–120 V AC± 10%, 50/60 Hz

220–240 V AC± 10%, 50/60 Hz