Model 409 High Temperature Controller Connections

Model 409.83 Temperature Controller

Installation

23

J4, J5 heating element

outputs

Socket-style connectors J4 and J5 transfer the heater control signal (line voltage)
from the temperature controller to the furnace heater elements. The cables
connected to J4 and J5 are interchangeable.

J4 and J5 Heating Element Output Connections

J6 serial

communication

The J6 serial communication port has three functions:

Interlock input

The interlock input can be connected to an external (normally-closed)
switch or contact. If the external switch or contact opens, the temperature
controller alarm is triggered, and power is cut off to the heating elements.

Interlock contact output

The interlock output (relay output) can be connected to an external
monitoring device to follow the enable/alarm state of the temperature
controller. This contact is closed when the Model 409.83 Controller heating
elements are enabled. If an alarm is triggered or the Disable button is
pressed (interlock is open), a signal is sent to the external monitoring device.

RS–485 communication port to temperature control module

All functions and readout values in each Eurotherm 2216 Controller can be
monitored and adjusted by a single RS-485 serial port located at J6.

If nothing is connected to the J6 port, a jumper plug must be installed (MTS part
number 525099-01).