Parameter settings, Parameter settings 30, Al list – MTS Model 409.83 Temperature Controller User Manual

Page 30

Advertising
background image

Model 409.83 Temperature Controller

30

Parameter Settings

Operation

Two seconds after releasing either button, the display blinks to show the
control module has accepted the new value.

2. Repeat step 1 for each control module.

You can return to the Home display by pressing the scroll and page buttons
simultaneously. Alternatively, you will always be returned to the Home
display if no button is pressed for 45 seconds.

3. Press the Enable Heater Elements switch.

This will bring the furnace up to temperature.

Parameter Settings

Note

Parameters are set at MTS Systems according to the configuration
drawing at the end of the manual. If a control module is replaced, set the
parameters according to the MTS-factory settings and then run the
Eurotherm Automatic Tuning. Once the MTS factory settings have been
entered, you may deviate from them (in small steps).

Parameters are settings within the control module that determine how the control
module will operate. For example, alarm setpoints are parameters that determine
when alarms will occur. Parameters are arranged in lists. The names of these lists
are found in the list headers. The available lists are:

Home List

Set Point List

On/Off List

Alarm List

Input List

Communications List

PID List

Output List

Access List

Typical List Header Display

Note

Refer to the Eurotherm Model 2216 product manual shipped with this
manual for a listing of the parameters available under each list header.

To change the value of a parameter:

1. From the Home display, press the page button to select the appropriate list

header.

2216

OP1

OP2

SP2

REM

AL
LiSt

List header

Always displays
‘LiSt’