MTS Series 311 Load Frame User Manual
Page 64
Advertising
Series 311 Load Frame
64
Aligning the Force Transducer
Maintenance
B.
Tighten to 5% of the torque recorded on the identification plate using
the appropriate bolt torque sequence.
Bolt Torque Sequence
C.
Rotate the indicator to see if the TIR is still 0.038 mm (0.0015 in) or
less. If not, loosen the nut or preloader collar and return to Step 3.
(Loosen the preloader collar following the sequence shown below.)
D.
Repeat Steps
and
to tighten the force transducer for the following
torque progression: 50%, 75%, and 100%.
Advertising