Go to step 5 – MTS Series 311 Load Frame User Manual

Page 65

Advertising
background image

Aligning the Force Transducer

Series 311 Load Frame

Maintenance

65

5. Finish the procedure.

In this step, you complete the force transducer alignment procedure.

A.

Remove the dial indicator.

B.

If you reduced pressure at the hydraulic power unit, restore full
hydraulic pressure.

C.

Turn the load frame’s hydraulic pressure to off.

Cap Screw Torque Specifications

load cell

*

*

Cap screw specifications for the Model 311.61 and 311.71 Load Frames
are contained on the drawings and parts lists for the specific unit.

Cap Screw Size

Torque

311.11

† with a 100 kN (22 kip) force transducer

5/8-11

70 N

m (52 lb

ft)

311.11

‡ with a 250 kN (55 kip) force transducer

5/8-11

176 N

m (130 lb

ft)

311.21)

1-8 UNC-2A

570 N

m (420 lb

ft)

311.31

1-8 UNC-2A

635 N

m (470 lb

ft)

311.41

§

§ with a 1500 kN (330 kip) force transducer

1-1/2-6

1220 N

m (900 lb

ft)

311.41

#

# with a 2500 kN (550 kip) force transducer

1-1/2-6

1760 N

m (1300 lb

ft)

311.51

1-1/2-6 UNC-2A

1490 N

m (1100 lb

ft)

Advertising