Tixfer - low-profile transducer interface transfer – MTS SWIFT 20 Ultra Sensor User Manual

Page 48

Advertising
background image

SWIFT 20 Sensors

48

TIXFER - Low-Profile Transducer Interface Transfer

Software Utilities

TIXFER - Low-Profile Transducer Interface Transfer

This program is used to change settings within the SWIFT Low Profile
Transducer Interface (TI). It can be used to read the current settings and save
them to the computer (upload) or write the settings to a Low-Profile TI with
values from a file on the computer (download).

Syntax

The tixfer command requires one argument:

tixfer port

port is the serial port number

1 = COM1:
2 = COM2:

Example

File Format

The file used with commands contains a header and a list of parameters.
Transducer calibrations can be uploaded and saved individually. If a test needs to
be rerun at a later date but the original transducer is not available, another
transducer can be used by downloading its calibration information.

The header must be the first thing in the file, and it must be in the form:

[SWIFT]

Following the header is a list of parameter settings. The syntax is:

ParamName=ParamValue

C:\bin>tixfer 1

tixfer $Revision: 1.11 $ ($Date: 1999/04/21 10:12:40 $)

0...exit

1...Upload/Save a chassis

2...Restore/Download a chassis

Enter choice? 1

Enter TI box address? 1

Enter output specifications filename? example.cal

Enter description? example for uploading settings

Allocating box...

Initializing box...

Advertising