Shunt error status, Shunt error status 58, Shunt error – MTS SWIFT 20 Ultra Sensor User Manual

Page 58: Status

Advertising
background image

SWIFT 20 Sensors

58

Shunt Error Status

Software Utilities

Shunt Error Status

This is the error status for each bridge from the last shunt cal that was performed.
The error status is a bit-mapped status word, where the error is present if the bit is
set to 1.

Bit 0 = Bridge amplifier was saturated prior to applying the shunt.

Bit 1 = Bridge amplifier was saturated after applying the shunt.

Bit 2 = Measured delta value failed high.

Bit 3 = Measured delta value failed low.

Bit 4 = A procedural error occurred (unable to read a value, etc.).

In the following example the error status received is 5 indicating that bits 0 and 2
were set.

Bit

4

3

2

1

0

Bit Value

16

8

4

2

1

Example

0

0

1

0

1

Error Status=

5

Advertising