Calibration File Elements

The following figure shows some elements of the calibration file:

Typical Calibration File

Items you may edit

OutputPolarities—defines the polarities of the six outputs. Change these

only if your application requires different polarities from those identified on
the transducer label.

AngleMode—selects the mode used for determining the encoder sine and

cosine.

AngleFixed—used for non-spinning applications.

AngleOffset— used for spinning applications. Normally you do not need to

change this value.

EncoderSize–defines the size of the encoder.

S10MC-33

SerialNum=10237221A
AngleMode=0
AngleOffset=0
AngleFixed=0
EncoderSize=2048
FXPolarity=0
FYPolarity=0
FZPolarity=0
MXPolarity=1
MYPolarity=0
MZPolarity=1

ZFX1=0
ZFX2=0
ZFY1=0
ZFY2=0
ZFY3=0
ZFY4=0
ZFZ1=0
ZFZ2=0

FX1Channel=2
FX2Channel=6
FY1Channel=99
FY2Channel=99
FY3Channel=99
FY4Channel=99
FZ1Channel=4
FZ2Channel=0
InputRange=10.0

KFX1=0.492315
KFX2=0.492315
KFY1=0.0
KFY2=0.0
KFY3=0.0
KFY4=0.0
KFZ1=0.49212
KFZ2=0.49212
KMX=0.0
KMYX=1.44278
KMYZ=1.44278
KMZ=0.0
KFXFY=0.0

Polarity for each output
(0 = normal, 1 = inverted)

Angle Mode (0 = Spinning, 1 = Fixed)

Angle Offset

Fixed Angle

Bridge Zeroes

Do Not Modify

Calibration Gains

Do Not Modify

Encoder Size