Calibration file elements, Calibration file elements 51, Items you may edit – MTS SWIFT 10 MC Sensor User Manual

Page 51

Advertising
background image

Select a Zero Method

SWIFT 10 MC Sensors

Setting up the Transducer Interface

51

Calibration File Elements

The following figure shows some elements of the calibration file:

Typical Calibration File

Items you may edit

•

OutputPolarities—defines the polarities of the six outputs. Change these

only if your application requires different polarities from those identified on
the transducer label.

•

AngleMode—selects the mode used for determining the encoder sine and

cosine.

•

AngleFixed—used for non-spinning applications.

•

AngleOffset— used for spinning applications. Normally you do not need to

change this value.

•

EncoderSize–defines the size of the encoder.

S10MC-33

SerialNum=10237221A
AngleMode=0
AngleOffset=0
AngleFixed=0
EncoderSize=2048
FXPolarity=0
FYPolarity=0
FZPolarity=0
MXPolarity=1
MYPolarity=0
MZPolarity=1

ZFX1=0
ZFX2=0
ZFY1=0
ZFY2=0
ZFY3=0
ZFY4=0
ZFZ1=0
ZFZ2=0

FX1Channel=2
FX2Channel=6
FY1Channel=99
FY2Channel=99
FY3Channel=99
FY4Channel=99
FZ1Channel=4
FZ2Channel=0
InputRange=10.0

KFX1=0.492315
KFX2=0.492315
KFY1=0.0
KFY2=0.0
KFY3=0.0
KFY4=0.0
KFZ1=0.49212
KFZ2=0.49212
KMX=0.0
KMYX=1.44278
KMYZ=1.44278
KMZ=0.0
KFXFY=0.0

Polarity for each output
(0 = normal, 1 = inverted)

Angle Mode (0 = Spinning, 1 = Fixed)

Angle Offset

Fixed Angle

Bridge Zeroes

Do Not Modify

Calibration Gains

Do Not Modify

Encoder Size