Troubleshooting – MTS Criterion Series 60 User Manual

Page 130

Advertising
background image

Troubleshooting

The following table describes some basic troubleshooting issues and possible fixes.

Reason

Check

Issue

Any pressed Emergency Stop will
prevent interlocks from being cleared.

Is an Emergency Stop button
pressed?

Cannot clear interlocks,
no apparent source for
the interlock.

Check if an E-Stop is pressed in the
fault status screen in TW.

The pump must be on to clear
interlocks.

Is the pump on?

Any fault that is detected will prevent
machine operation. Generally, contact

Is there a fault on either the Interface
indicator on the controller box or MTS

MTS technical support to troubleshoot.
Faults are reset when power is cycled.

Criterion Series 60 logic indicator on
the IOP panel? Check the indicator
LEDs or check the software fault
status.

The guard connector contains jumpers
to configure the guard as either “slow”

Is something plugged into the guard
connector in the J11 connector? Make

or “interlock”. When nothing is plugged

sure the guard connection is not

in, the guard gets configured as an
interlock and the guard appears open.

plugged into the adjacent auxiliary
power connector.

Although the interlock connection on
the MTS Insight board is rarely used,

Is something plugged into the J3
connector on the MTS Insight
controller board?

if a jumper plug is not installed, an
interlock will be generated.

The guard function “slow” or “interlock”
is configured by jumpers in the cable.

Is the guard door open and the guard
configured to Interlock?

If no jumpers are installed, it is
configured to interlock when the guard
door is open.

An upper limit can be tripped if the
piston extends too far (risk of

Check the upper actuator limit.

An interlock is active.

bypassing seal), or if the piston is too
close to the crosshead (piston moving
up or crosshead moving down).

A lower limit is tripped if the upper and
lower grips in the upper test space are

Check the lower actuator limit.

130 | MTS Criterion

Series 60

Troubleshooting

Advertising