MTS Landmark Tabletop User Manual

Page 97

Advertising
background image

Checking the Precharge Pressure

Description

Item

Locknut

1

Valve Stem

2

Close

3

Open

4

Poppet Type Accumulator Valve. To open and close, use locknut.

5

Or

6

Valve Core

7

Core type accumulator valve. To open and close use chuck valve.

8

THESE ITEMS SUPPLIED BY USER

9

Nitrogen Bottle Valve

10

Nitrogen Bottle

11

Nitrogen Bottle Pressure Gage

12

Regulator Output Pressure Gage

13

Regulator Shut-Off Valve

14

Regulator Output Pressure Valve

15

Nitrogen Supply Hose

16

Input Check Valve

17

Bleed Valve

18

Extension Hose

19

Gage Protector (factory set to limit pressure to the gage to approximately 1.4 MPa
(200 psi)

20

High Pressure Gage—0–21 MPa (0–3000 psi)

21

Low Pressure Gage—0–2.1 MPa (0–300 psi)

22

Chuck Valve

23

2. With an open-end wrench, turn the locknut counterclockwise on the accumulator valve assembly to open

the valve. Read the pressure on either the high or low accumulator charging kit pressure gage.

a) If the pressure reading is other than the required pressure level recorded on the accumulator, continue

with the next subsection, “Change the Precharge Pressure”.