Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

8

09/18/14

The “Volume Dispensed” accumulations, for infusion and withdrawal, are reset to 0 when:

A) The pump is powered on.
B) The syringe diameter is changed.
C) From the RS-232 clear “Volume Dispensed” command (CLD) or Clear Dispense function.
D) The accumulated Volume Dispensed rolls over from 9999 to 0.
E) A sub-program is selected when the Pumping Program executes a Sub-Program Selection

function.

5.8 Resuming When Paused

If the Pumping Program is stopped before the completion of the Pumping Program, the response prompt will
change to 'P', indicating that the Pumping Program is paused. While paused, starting the pump again will
resume the Pumping Program where it was stopped. This means that the Pumping Program will continue at the
point in the Phase where it was stopped and the ‘Volume to be Dispensed’ will still be referenced from when
the Program Phase first started.

Pause is exited by sending a stop command "STP", or changing any setting. When the Pumping Program is
started again, it will start from the beginning (Phase 1).

5.9 Pump Stalled

Model: NE-501 only

When the operation of the motor is impeded due to excessive force needed to drive the syringe, the pump will
stop, pausing the Pumping Program, and a stall alarm will occur.

The buzzer will sound continuously if alarms are enabled. Also, if the RS-232 Safe Mode is enabled, an auto-
alarm message will be sent to an attached computer.

Sending any command will stop the buzzer and clear the alarm. When the problem causing the pump motor to
stall has been corrected, the Pumping Program can be resumed from any start trigger source: TTL input, or RS-
232.

NOTE: Pump stall is an error condition and should not be used as a normal method of stopping the pump.
Continued pump stalling can damage the drive nut block. Damage to the drive nut block is not covered by the
warranty.

6. Setup Configuration

The following configuration parameters can be set or queried:

The values for "n" are "0" for off, "1" for on. See relevant sections for command definitions.

Command

Description

ADR nn

Set the pump network address or mode of communications.

PF n

Power Failure mode, where ‘n’ is the current setting.

LN n

Low noise mode to minimize the motor’s micro-stepping resonance

AL n

Alarm mode, where ‘n’ is the current setting.

TRG aa

Operational Trigger default configuration, where ‘aa’ is the current configuration.

DIN n

Directional control TTL logic input setup

ROM n

Pump motor operating TTL logic output configuration

LOC n

Lock out mode. Sets Auto-Run mode with Sub-Programs

BP n

Notification beep mode, where ‘n’ is the current setting.