New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 36

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
31
09/18/14
Set:
Currently selected Phase is set to <phase data>. Previous Phase is stored in non-volatile memory and
the requested Phase is recalled from the non-volatile memory. Set is only valid if the Pumping
Program is not operating.
Query response:
<phase data>
Currently selected Phase
PUMPING PROGRAM FUNCTION
FUN [ <phase function> ]
Set/query the Pumping Program Phase’s function.
This command is relevant to the currently selected Phase. Set is only valid if the Pumping Program is
not operating
For a more detailed description of Program commands, see sec. 7.3, Program Function Descriptions.
<phase function> =>
Rate Data Functions
When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and ‘DIR’
commands to setup the pumping parameters for the current phase.
RAT
Pumping rate.
FIL
Fill syringe to dispensed volume
INC
Increment rate.
DEC
Decrement rate.
Non-Rate Control Functions
STP
Stop pump.
PAS <number data>
Pauses pumping for ‘nn’ seconds, or until trigger if 0.
PAS <n.n>
Pauses pumping for 'n.n' seconds.
LOP <count data>
Loop to previous loop start ‘nn’ times.
LPS
Loop starting Phase.
LPE
Loop end Phase.
JMP <phase data>
Jump to Program Phase.
TRG <nn>
Override Operational Trigger Default Mode, set to <nn>
Conditional Functions
IF
<phase data>
If Program input TTL pin low, jump to Phase.
EVN <phase data>
Set event trigger trap.
EVS <phase data>
Set event square wave trigger trap.
EPL <expansion pin>
Pause and wait for expansion pin event edge or level.
EPE <expansion pin>
Pause and wait for expansion pin event edge.
EVE <expansion pin>
Re-direct event trigger trap to expansion pin.
EVR
Event trigger trap reset.
CLD
Clear total dispense volume.
TTL Logic and Other Functions
OUT <TTL level>
Set programmable output pin.
OE0 <expansion pin>
Set expansion pin to 0 (low level).
OE1 <expansion pin>
Set expansion pin to 1 (high level).
BEP
Sound short beep.
PUMPING RATE
RAT [ C | I ] [ <float> [ <rate units> ] ]
Set/query pumping rate.
<rate units> => UM
=
µL/min
MM
= mL/min
UH
=
µL/hr
MH
= mL/hr