3 example 3: ramping the flow rate – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 24

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

19

09/18/14

When entering a function with associated data, such as with the ‘Pause’ in Phase 5 or the ‘Loop’ in Phase 6, the
function is entered in 2 steps. First select the function and store it. Then enter the associated data.

Phase

Function

Rate

Volume

Direction

1

RAT

750 mL/hr

2.0 mL

Infuse

Phase

Function

Rate

Volume

Direction

2

RAT

750 mL/hr

0.25 mL

Withdraw

Phase

Function

3

LPS

Phase

Function

4

LPS

Phase

Function

5

PAS 90

Phase

Function

6

LOP 03

Phase

Function

7

BEP

Phase

Function

8

PS:30

Phase

Function

Rate

Volume

Direction

9

RAT

750 mL/hr

2.25 mL

Infuse

Phase

Function

Rate

Volume

Direction

10

RAT

750 mL/hr

0.25 mL

Withdraw

Phase

Function

11

LPE

7.4.3

Example 3: Ramping the Flow Rate

m

l/

h

r

R

a

te

Time

0

150

200

250

Phases
2 - 4

Phases
5 - 8

Phases
9-11

Note: Graph is representative only.
Pumping rate increments and decrements in steps.

Continuously ramp up and down the pumping rate. Starting at 200 mL/hr, the pumping rate will increment to
250 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Then the pumping rate will decrement to
150 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Finally, the pumping rate is incremented
back to 200 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed, then the process is repeated.

Phase

Function

Rate

Volume

Direction

1

RAT

200 mL/hr

0.1 mL

Infuse

Phase

Function

2

LPS

Phase

Function

Rate

Volume

Direction

3

INC

1.0

0.1 mL

Infuse