Appendix, 1 rs-232 command summary – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual
Page 46

New Era Pump Systems Inc.
www.SyringePump.com
Model NE-500/NE-501 Multi-Phaser™
Publication #1200-02
41
09/18/14
10. Appendix
10.1 RS-232 Command Summary
<command> =>
DIA [ < float > ]
Syringe inside diameter
PHN [ < phase data > ]
Program Phase number
FUN [ < phase function > ]
Program Phase function
< phase function > =>
RAT
Pumping rate.
FIL
Fill syringe to volume dispensed.
INC
Increment rate.
DEC
Decrement rate.
STP
Stop pump.
PAS <number data>
Pauses pumping for ‘nn’ seconds, if 0, wait for start trigger.
PAS [n.n]
Pauses pumping for 'n.n' seconds.
JMP <phase data>
Jump to Program Phase.
PRI
Sub-Program Selection Input from expansion connector
PRL <number data>
Sub-Program Start Label definition.
LPS
Loop starting Phase.
LOP <count data>
Loop to previous loop start ‘nn’ times.
LPE
Loop end Phase.
IF <phase data>
If Program input low, jump to Program Phase.
EVN <phase data>
Set event trigger.
EVS <phase data>
Set event square wave trigger.
EPL <expansion pin>
Pause program, wait for low edge or level on expansion pin
EPE <expansion pin>
Pause program, wait for falling edge on expansion pin
EVE <expansion pin>
Re-direct event trigger to expansion pin
EVR
Event trigger reset.
CLD
Clear total dispense volume.
OUT { 0 | 1 }
Set programmable output pin.
OE0 <expansion pin>
Set expansion pin low
OE1 <expansion pin>
Set expansion pin high
TRG <nn>
Override Operational Trigger default setting
BEP
Sound short beep.
RAT {C | I ] [ <float> [ UM | MM | UH | MH ]] Pumping rate
VOL [ <float> | <volume units>]
Volume to be Dispensed, or set Volume units
DIR [ INF | WDR | REV | STK ]
Pumping direction
RUN [ <phase data> ]
Starts the Pumping Program
[E [<phase data>] ]
Pumping Program event trigger
PUR
Start purge
STP
Stop/pauses the Pumping Program
DIS
Query volume dispensed
CLD { INF | WDR }
Clear volume dispensed
SAF [ <n> [ <n> [ <n> ] ] ]
Safe communications mode
LN [ 0 | 1]
Low motor noise mode
AL [ 0 | 1]
Alarm buzzer mode
PF [ 0 | 1]
Power failure auto-restart mode
TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2 | RL | RH | SL | SH | OF ]
Operational trigger default mode
DIN [ 0 | 1]
Directional input control mode
ROM [0 | 1]
Pump Motor Operating TTL output mode
LOC [ P ] [ 0 | 1]]
Keypad lockout mode or Program Entry Mode lockout
BP [ 0 | 1]
Key and Notification beep mode
OUT [ E <expansion pin> | 5] { 0 | 1 }
Set TTL output level
IN { E <expansion pin> | 2 | 3 | 4 | 6 }
Query TTL input level
BUZ [ 0 | { 1 [ < n > ] } ]
Buzzer control
VER
Query firmware version
System Commands: Valid regardless of current network address
*ADR [ <n> [<n>] [ B <baud-rate>] ]
Network address and baud rate