1 ttl i/o operational controls, Db-9 connector ttl i/o – New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 43

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

38

09/18/14

DB-9 Connector TTL I/O

Pin #

Definition

Type

Function

1

Vcc (5V)

Reference

Logic high reference. Power on indicator.

2

Operational Trigger

Command: TRG [aa]

Input

Configurable start/stop operational trigger input.
FT

Foot Switch

Falling edge: Start or stop trigger

FH

Foot Switch Hold

Falling edge: Start trigger

Rising edge: Stop trigger

F2

Foot Switch Reverse Rising edge: Start or stop trigger

LE

Level

Falling edge: Stop trigger

Rising edge: Start trigger

ST

Start only

Falling edge: Start trigger

T2

Start only Reverse

Rising edge: Start trigger

SP

Stop only

Falling edge: Stop trigger

P2

Stop only Reverse

Rising edge: Stop trigger

RL

Start on low level

Low level:

Start trigger

RH

Start on high level

High level:

Start trigger

SL

Stop on low level

Low level:

Stop trigger

SH

Stop on high level

High level:

Stop trigger

ET

Program function: redirect trigger to Event trap

BT

Program function: Redirects ‘Stop’ key to Event trap

3

Pumping Direction

Input

Changes pumping direction according to setup
Command:

DIN 0

DIN 1

Falling edge: Infuse

Withdraw

Rising edge: Withdraw

Infuse

4

Event Trigger

Input

Event input or user definable input

5

Program Output

Output

Program controlled output or user definable output

6

Program Input

Input

Program conditional input read by the “IF” program function.
Also user definable input.

Also used by the keypad lockout function.

7

Pump Motor Operating

Output

Setup Command: ROM 0
High: Pumping; Low: Not pumping

Setup Command: ROM 1
High: Pumping or Pause timer; Low: Not pumping

8

Pumping Direction

Output

High: Infuse; Low: Withdraw

9

Ground (0V)

Reference

Logic low reference

9.1 TTL I/O Operational Controls

Operational Trigger (Pin 2): The input signal on this pin controls the operation of the Pumping Program. Its
functionality is user configurable. Use the ‘TRG' command to configure the default settings for this input pin
(See Section 6.5, TTL I/O Operational Trigger Default Configuration).

Each option defines when the Operational Trigger input is activated. When activated, the trigger emulates a
‘Start/Stop’ key.

Foot Switch:

Operates like a ‘Start/Stop’ key, whereby each falling edge (contact to ground) either starts
or stops/pauses the Pumping Program.

Foot Switch Hold: Falling edge starts the Pumping Program and the rising edge stops the Pumping Program.

With a foot switch, the Pumping Program will run while the foot switch was held down.