New Era NE-500 Multi-Phaser Programmable OEM Syringe User Manual

Page 29

Advertising
background image

New Era Pump Systems Inc.

www.SyringePump.com

Model NE-500/NE-501 Multi-Phaser™

Publication #1200-02

24

09/18/14

Phase

Function

2

LPS

Phase

Function

3

PRI

Phase

Function

4

PRL 01

Phase

Function

Rate

Volume

Direction

5

RAT

100.0 mL/hr

10 mL

Infuse

Phase

Function

6

JMP 12

Phase

Function

7

PRL 02

Phase

Function

Rate

Volume

Direction

8

RAT

500.0 mL/hr

10 mL

Infuse

Phase

Function

9

JMP 12

Phase

Function

10

PRL 03

Phase

Function

Rate

Volume

Direction

11

RAT

750.0 mL/hr

10 mL

Infuse

Phase

Function

12

LOP 05

Phase

Function

13

JMP 01

7.4.8

Example 8: Dispensing with Complex Synchronization

:01

Repeat if
Program Input
low

:10

:10

Restart on Event Trigger
or 10 seconds

Event Trigger

W

it

h

d

ra

w

I

n

fu

s

e

TTL Program Ouput

TTL High

TTL Low

Event Trigger

Infuse until
Event Trigger

This example demonstrates a complex interaction with external equipment, such as synchronizing with another
syringe pump. The Program includes a variety of interactions with external equipment, which demonstrates the
various control possibilities of the NE-500/501.

The Pumping Program begins by canceling any previous event traps (PHASE 01) and raising the ‘Program

Advertising