Example 4: Complex Dispenses with External Synchronization

0

0

750

W

it

h

d

ra

w

(

m

l/

h

r)

I

n

fu

s

e

Start
Trigger

Start
Trigger

:60

:60

:60

Start
Trigger

Start over

Syringe

Refill

A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during
the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for
a start trigger before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
dispenses, the syringe is refilled by the amount infused, 17.25 mL. Then the buzzer beeps, to alert the operator
to the start of the first set of dispenses. The process is then repeated.

Phase

Function

Rate

Volume

Direction

1

RAT

750.0 mL/hr

0.5 mL

Infuse

Phase

Function

Rate

Volume

Direction

2

RAT

300.0 mL/hr

1.5 mL

Infuse

Phase

Function

3

BEP

Phase

Function

4

PAS 00