10

09/18/14

Condition

Buzzer Action

Alarm condition, such as pump motor stalled

Steady alarm

Any command sent to the pump will stop the alarm.

6.5 TTL I/O Operational Trigger Default Configuration

Command: "TRG <aa>", <aa> as below in Settings.

Configures the default mode of the TTL I/O ‘Operational Trigger’ (pin 2) that will control the Pumping
Program’s operation. This default setting can be overridden within a Pumping Program. (See sec. 9.1, TTL I/O
Operational Controls

). The 2 letter configuration Setting parameter <aa> is defined in column ‘Setting’ as

follows:

Setting

<aa>

RS-232 Command Setting

FUN TRG <n>

<n> =>

Name

Function

FT

0

Foot Switch

Falling edge starts or stops the Pumping Program

FH

1

Foot Switch Hold

Falling edge starts the Pumping Program
Rising edge stops the Pumping Program

F2

2

Foot Switch
Reversed

Rising edge starts or stops the Pumping Program

LE

3

Level Control

Falling edge stops the Pumping Program
Rising edge starts the Pumping Program

ST

4

Start Only

Falling edge starts the Pumping Program

T2

5

Start Only Reversed

Rising edge starts the Pumping Program

SP

6

Stop Only

Falling edge stops the Pumping Program

P2

7

Stop Only Reversed

Rising edge stops the Pumping Program

RL

8

Start on Low Level

Start Pumping Program on a low level

RH

9

Start on High Level

Start Pumping Program on a high level

SL

10

Stop on Low Level

Stop Pumping Program on a low level

SH

11

Stop on High Level

Stop Pumping Program on a high level

OF

12

Trigger off

Disable trigger

Et

13

Trigger Event

Pumping Program Function Only: Stop Trigger
triggers an event instead of stopping the pump

Bt

14

Button Trigger Event

Pumping Program Function Only: Pressing ‘Stop’
key triggers an event instead of stopping the pump

6.6 TTL I/O Directional Control Input Configuration

Command: "DIN <n>"

Configures how the TTL input ‘Pumping Direction’ (pin 3) will control the pumping direction. (See sec. 9.1,
TTL I/O Operational Controls

).

Setting <n>

Mode

Function

0

Reciprocating Pumps

Falling edge: Infuse; Rising edge: Withdraw
- Used for 2 pumps operating in opposite directions

1

Dual Pump

Falling edge: Withdraw; Rising edge: Infuse
- Used with 2 pumps operating in the same direction