To accept the alarm logic setting and continue the program, press the ENTER button. The
program displays the Alarm Silence screen.

-

To update the alarm logic setting:
1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program. The program displays the Alarm Silence
screen.

Alarm Silence Screen
This screen indicates how the buzzer operates during alarm conditions. If the setting is

ON

, you can

use the reset switch to silence the buzzer during alarm conditions. If the setting is

O F F

, you cannot

silence the buzzer until all alarm conditions pass. The default setting is

O N

.

-

To accept the alarm silence setting and continue the program, press the ENTER button. The
program displays the Alarm Hysteresis screen.

-

To update the alarm silence setting:
1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program. The program displays the Alarm
Hysteresis screen.

Alarm Hysteresis Screen
This screen indicates when the PioneerR-16 considers an alarm condition passed. The alarm
hysteresis feature helps prevent readings that are near an alarm setpoint from frequently going into
and out of alarm (relay chattering). For example, if the alarm hysteresis setting is 2.0% of fullscale,
fullscale is 10.0 ppm, and the alarm 1 setpoint is 3.0 ppm; then the Pioneer will not consider the alarm
1 condition “passed” until the reading goes below 2.8 ppm. You can choose from a range of 0.0% to
10.0% of fullscale (in 0.1% increments). The default setting is 2.0%.

-

To accept the alarm hysteresis setting and continue the program, press the ENTER button. The
program displays the Calibration Time Out screen.

-

To update the alarm hysteresis setting:
1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program. The program displays the Calibration Time
Out screen.

Calibration Time Out Screen
The Pioneer-R16’s Calibrate program includes two separate time-out features. The first time-out
feature is a general 5-minute time-out. The Pioneer automatically returns to the normal screen if you
do not move from one Calibrate screen to another within 5 minutes. You cannot adjust this setting,
and it affects all Calibrate screens except the Fresh Air and Calibration screens.

Pioneer-R16 Operator’s Manual