RKI Instruments Pioneer 16R User Manual

Page 20

Advertising
background image

instrument, display, and channel software programs.
1.

Press and hold down the reset switch to enter the program.

2.

When the DISPLAY VERSIONS screen appears, release the reset switch.

3.

Press and release the reset switch to display the remaining software versions and return to the
normal screen.

To display TWA and STEL readings:
The Pioneer continuously computes Time Weighted Average (TWA) and Short Term Exposure Limit
(STEL) readings for all active channels.

N O T E

: The Pioneer does not calculate or display TWA and STEL setting for oxygen channels or

combustible gas channels that have a unit of measure setting of %LEL. However, the Pioneer does
calculate and display TWA and STEL settings for combustible gas channels that have a unit of
measure setting of PPM/LEL.

The STEL reading represents a channel’s average gas readings for the past 15 minutes.
The TWA reading represents a channel’s average gas readings for the past 8 hours.
1.

Press and hold down the reset switch to enter the program.

2.

When the DISPLAY TWA/STEL screen appears on the display screen, release the reset switch.
The STEL reading displays for the first active channel.

3.

Press and release the reset switch to display the TWA reading for the first active channel.

4.

Continue pressing and releasing the reset switch to display the STEL and TWA readings for the
remaining channels.

5.

Press and release the reset switch to return to the normal screen.

To display minimum and maximum readings:
The Pioneer stores the minimum and maximum reading for all active channels since the last time the
minimum and maximum readings were reset.
1.

Press and hold down the reset switch to enter the program.

2.

When the DISPLAY MIN/MAX screen appears on the display screen, release the reset switch.
The minimum reading displays for the first active channel.

3.

Press and release the reset switch to display the maximum reading for the first active channel.

4.

Continue pressing and releasing the reset switch to display the minimum and maximum readings
for the remaining channels.

5.

Press and release the reset switch to return to the normal screen.

To display alarm sets:
The Display Setpoints and Readings program displays the alarm sets for all active channels. The alarm
sets include the alarm logic, alarm silence, alarm setpoint, and alarm type settings.

N O T E

:

You can change the alarm logic and alarm silence settings in the Instrument Setup
program. You can change the alarm setpoint and alarm type settings in the Channel Setup
program.

1.

Press and hold down the reset switch to enter the program.

2.

When the ALARM SETS screen appears on the display screen, release the reset switch. The
alarm logic setting (LATCHING or AUTO-RESET) displays for all active channels.

3.

Press and release the reset switch to display the alarm silence setting (ON or OFF) for all active
channels.

4.

Continue pressing and releasing the reset switch to display the setpoint and alarm type for the fail,
alarm 1, alarm 2, and alarm 3 alarms of channel 1.

5.

Repeat step 4 to display the setpoint and alarm type for the remaining active channels.

6.

Press and release the reset switch to return to the normal screen.

To perform the self check:

The Pioneer includes a self check feature that makes sure the analyzer card(s) and microprocessor are
operating correctly.

Pioneer-R16 Operator’s Manual

Operation

19

Advertising