The second time-out feature affects the Fresh Air and Calibration screens only. These are the screens
you use during calibration to adjust the zero and span readings (either at the transmitter or the
Pioneer). The Pioneer automatically returns to the normal screen if you do not move from the Fresh
Air or Calibration screen within the time-out setting specified in the screen below. You can choose
from a range of 10 to 100 minutes (in increments of 10). The default setting is 10 Minutes.

-

To accept the calibration time out setting and continue the program, press the ENTER button.
The program displays the Instrument ID screen.

-

To update the calibration time out setting:
1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program. The program displays the Instrument ID
screen.

Instrument ID screen
This screen indicates the ID number (address) for this Pioneer-R16. The instrument ID is recorded in
the datalog when the datalog records instrument readings. You can choose from a range of 1 to 255.
The default setting is 1.

-

To accept the instrument ID setting and continue the program, press the ENTER button. The
program displays the Noise Filter screen.

-

To update the instrument ID setting:
1.

Use the UP/YES or DOWN/NO button to display the setting you want.

2.

Press the ENTER button to continue the program. The program displays the Noise Filter
screen.

Noise filter screen
This screen indicates the length of time that the Pioneer-R16 averages data. You can choose from a
range of 0.5 to 15.0 seconds (in increments of 0.5). The default setting is 2.5 Seconds.

N O T E

:

The Pioneer-R16 continuously displays gas readings although it averages gas readings

according to the noise filter setting.

Pioneer-R16 Operator’s Manual

Operation

16