Care and use manual – Waters XBridge Protein BEH, C4, 300A, 3.5 µm Columns User Manual
Page 4
 
4
[ CARE AND USE MANUAL ]
XBridge Protein BEH C
4
, 300Å, 3.5, 5, 10 �m
Mobile Phase:
A: 0.1% TFA in water
B: 0.075% in 71.4% acetonitrile/28.6% water
Gradient:
Time (Column Length)
50 mm
100 mm
150 mm
250 mm
%A
%B
Curve
Initial
Initial
Initial
Initial
72
28
*
25
50
75
125
0
100
6
27
54
81
135
0
100
1
45
90
135
225
72
28
1
Temperature:
40 °C
Detection:
220 nm
1
2
3
4
5
6
1 - RNase A
2 - Cyt. C
3 - BSA
4 - Myoglobin
5 - Enolase
6 - Phosphorylase b
Figure 2: Typical Protein Test Mixture Chromatogram using MassPREP Protein 
Standard Mixture
This chromatogram is typical of the results obtained in Waters 
laboratories with the method described above, using a XBridge 
Protein BEH C
4
, 300Å, 3.5 µm, 2.1 x 50 mm Column. The
retention times will double, triple, and be five times greater 
for the 100 mm, 150 mm, and 250 mm columns respectively.
The exact results observed in any laboratory will depend on 
the instrument in use. System volume, gradient generation 
mechanism, mixing, design of temperature control, detector cell 
dimensions, detector optical properties, and detector electronic 
properties all have a direct impact on the observed chromatogram. 
The pattern should be similar, however, on any well-functioning, 
modern HPLC. This test is exceptionally valuable for monitoring 
the life of the column and for troubleshooting separation 
difficulties that may arise.
III. COLUMN USE
To ensure the continued high performance of XBridge Protein BEH C
4
,
300Å, 3.5, 5, and 10 µm Columns, follow these guidelines:
a. Sample Preparation
Sample impurities often contribute to column contamination. 
Samples should be free of particles before injection into the system. 
It is preferable to prepare the sample in gradient solvent A or in a 
mobile phase that is weaker (less organic modifier) than the initial 
strength mobile phase. This ensures the best peak shape and.
If the sample is not dissolved in the mobile phase, ensure that the 
sample, solvent and mobile phases are miscible in order to avoid 
sample and/or buffer precipitation.
Filter sample with 0.2 μm filters to remove particulates. If the 
sample is dissolved in a solvent that contains an organic modifier 
(e.g., acetonitrile, methanol, etc.) ensure that the filter material 
does not dissolve in the solvent. Contact the filter manufacturer 
with solvent compatibility questions. Alternatively, centrifugation 
for 20 minutes at 12–50,000 g, followed by the transfer of the 
supernatant liquid to an appropriate vial, could be considered.
b. Operating pH Limits
The recommended operating pH range for XBridge Protein BEH 
C
4
, 300Å, Columns is 1 to 12. A listing of commonly used buffers
and additives is given in Table 2. Additionally, the column 
lifetime will vary depending upon the operating temperature as 
well as the type and concentration of buffer used.